CAR-Colorimetric-96 tests – CAR (Clinical Array Reader)
The autoclart is a bench-top semi-automated system for processing microplate-based assays.
The autoclart is designed for microplate processing and provides reagent addition, liquid handling, heating, cooling, and shaking of the microplate.
The microplate well holder offers:
- heating and cooling under a bespoke peltier controlled system, giving precise temperature control during the assay
- and shaking of the wells during incubation
The autoclart contains an X/Y/Z mechanism for addition and aspiration of reagents, and incorporates a tip picking and ejection system for use with disposable tips. Precise volume control is achieved using a peristaltic pump, driven by micro stepping motors under software control. Incorporated into the tip picking system are 4 aspirating needles, for the removal of waste reagent, which are connected to the peristaltic pump. The minimum number of samples that can be processed during a run is 4 and the maximum is 96.
The autoclart operator interface is via an integral Control Panel containing a control knob and LCD display screen. The Control Panel allows the operator to select the number of samples to be processed and run the appropriate assay. The display screen provides visual feedback outlining the progress of the assay.
The base of the instrument contains positions for:
- a Tip Rack Holder
- Reagent Trough containing 3 vials
- a separate Pre-Heated Hybridisation Trough
- a Waste priming trough
- and Tip Waste Container
All the consumable waste is housed inside the instrument, giving the unit a small footprint in the lab.